V&F PEMSense

Manufacturer: V&F

V&F PEMSense

EI-MS analyzer for rapid measurement of H2, N2, O2, H2O and CO2 on PEM fuel cells, with a dual sample gas inlet for anode and cathode.

Key features

  • Dual sample gas inlet for anode and cathode
  • Cathode response time T90 < 150 ms
  • Very low anode gas consumption of 3–60 ml/min
  • Anode inlet pressure up to 6 bar(a)

Technical data

Measuring principle
Electron impact mass spectrometry (EI-MS), quadrupole
Mass range
1–100 amu
Resolution
< 1 amu
Analysis time
>= 32 ms
Measuring range
0–100 vol%
Response time (cathode)
T90 < 150 ms
Response time (anode)
T90 < 2 s @ 4 bar(g)
Detection limit
<= 100 ppm
Concentration drift
< ±5 % over 24 h
Accuracy
< ±3 %
Precision
< ±3 %
Ambient temperature
20–35 °C
Gas consumption (cathode)
3000 ml/min
Gas consumption (anode)
3–60 ml/min
Gas inlet temperature
50–180 °C, adjustable
Gas inlet pressure (cathode)
0.75–2 bar(a)
Gas inlet pressure (anode)
0.75–6 bar(a)
Power supply
230 V/50 Hz or 115 V/60 Hz, 800 W
Dimensions (W x H x D)
534 x 743 x 639 mm
Weight
62 kg

Description

The V&F PEMSense is an electron impact mass spectrometer (EI-MS) for the rapid measurement of H2, N2, O2, H2O and CO2 on PEM fuel cells.

The analyzer has a dual sample gas inlet: the anode side with a gas flow below 60 ml/min, and the cathode or exhaust side with a gas flow below 3000 ml/min.

Typical applications

  • Fuel cell testing
  • Fuel cell development

Application examples

Automotive industry

  • Gas analysis in fuel cells

Measurable compounds

  • Permanent gases: H2, O2, N2, CO, CO2
  • Moisture: H2O
  • Noble gases: Ar
  • Hydrocarbons: CH4

Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.