Manufacturer: V&F
V&F AirSense
Multi-purpose IMR-MS mass spectrometer for an unlimited number of gas components, built for flexibility and robustness across automotive, food, environmental and medical applications.
Key features
- Patented ion-molecule reaction (IMR-MS) technology
- Unlimited number of measurable gas components
- Mass range 7–519 amu with detection limits below 100 ppt
- Response time T90 < 150 ms for transient processes
- Heated gas inlet adjustable from 50 to 190 °C
Technical data
- Measuring principle
- Ion-molecule reaction mass spectrometry (IMR-MS)
- Mass range
- 7–519 amu
- Resolution
- < 1 amu
- Analysis time
- >= 10 ms
- Measuring range
- 10^5
- Response time
- T90 < 150 ms
- Detection limit
- < 100 ppt
- Concentration drift
- < ±5 % over 24 h
- Accuracy
- < ±3 %
- Precision
- < ±3 %
- Ambient temperature
- 20–35 °C
- Max. humidity
- 80 % (non-condensing)
- Gas consumption
- 30–3,000 ml/min
- Gas inlet temperature
- 50–190 °C
- Gas inlet pressure
- 0.75–2 bar(a)
- Power supply
- 230 V/50 Hz or 115 V/60 Hz, 1000 W
- Dimensions (W x H x D)
- 534 x 743 x 639 mm
- Weight
- 87 kg
Description
The V&F AirSense is a gas analyzer built around a mass spectrometer that uses the patented ion-molecule reaction (IMR-MS) principle. It is the multi-purpose model in the V&F range, intended for applications that involve an unlimited number of gas components and that call for maximum flexibility and robustness.
Typical applications
- Automotive applications
- Food, beverage and tobacco applications
- Environmental applications
- Medical applications
Application examples
Automotive industry
- Online oil dilution measurement
- Gas analysis in fuel cell technology
- Ammonia and NOx measurements in engine exhaust gas
- Desulfation strategies for catalytic converters
Environment – immission
- Immission measurements with IMR-MS according to EPA methods
- Immission measurement of gases near waste incineration plants
- Mobile air pollution measurement in traffic
- Air quality measurement near petroleum refineries
Environment – emission
- Online quantification of tar during biomass gasification
- Dioxin formation in waste incineration plants
Environment – ambient air quality
- Ambient air quality and workplace exposure limit measurements
- Real-time ambient air pollution analysis for photocatalytic research
Specialty and industrial gases
- CO2 raw gas measurement and process control
- Analysis of N2O, CO2 and hydrocarbons in pure O2
- Hydrocarbons and other impurities in calibration and specialty gases
Food, beverage and tobacco industries
- Rapid detection of impurities in raw milk
- Measurement of flavour release
- Fast detection of mould in jams
- Determination of nut quality
- Analysis and quantification of short-chain fatty acids in fermentation processes
- Quality control in PET recycling (PET flakes)
- Real-time measurement of styrene from polystyrene packaging
- Gas-phase analysis of cigarette paper aroma
- Determination of volatile compounds in cigarette packaging
- Puff-by-puff online analysis of cigarette smoke (Hoffmann analytes) in real time
Healthcare
- Non-invasive diagnosis of diseases by breath gas analysis
- Early detection of bacteria
- Monitoring of propofol concentration in exhaled air
- Bioavailability of administered drugs
- Halitosis after drug administration
Cement industry
- Detection of pollutants from cement production
- Emission measurements of alternative fuels in the cement industry
- Observation of benzene formation in cement kilns
Steel industry
- Process control in steel production
- Monitoring of hazardous exhaust gases in steel production
- Hydrogen production in a cyclic steelmaking process
Alternative energies
- Emission reduction using hydromethane
- Monitoring of gasification processes
- Gas analysis in fuel cell technology
Polymer, tyre and rubber industry
- Process monitoring in polymer production for the tyre industry
- Quality control in PET recycling (PET flakes)
- Real-time measurement of styrene from polystyrene packaging
- Evaporation of inorganic and organic compounds from elastomers
Measurable compounds
- Inorganic compounds: H2O, O2, CO2, CO, H2S, SO2, S, NH3, NO, NO2, N2O, HCl
- Aromatics: C6H6 (benzene), C7H8 (toluene), C8H8 (styrene), C8H10 (xylene/ethylbenzene), C6H6O (phenol), PAHs
- Alkanes: CH4, C2H6, C3H8, C4H10, C5H12, C6H14, C7H16, C8H18, C9H20, C10H22
- Alkenes: C2H4, C3H6, C4H8, C5H10, C6H12, C7H14, C8H16, C9H18, C10H20, C4H6 (1,3-butadiene)
- Alkynes: C2H2 (acetylene), C3H4 (propyne)
- Sulfur compounds: H2S, COS, SO2, CS2, CH3SH, C2H5SH
- Nitrogen compounds: NH3, NO, NO2, N2O
- Alcohols: CH3OH, C2H5OH, C3H7OH, C4H9OH
Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.
Ask about this machine
Send us your question or the details of your project. We usually reply within one working day.
More in this range

V&F CO2Sense
IMR-MS analyzer developed for measuring organic and inorganic impurities in CO2, meeting common food and beverage CO2 quality specifications such as ISBT and EIGA.
V&F
V&F PETSense
IMR-MS sniffer for the fast, accurate detection of contaminated refillable PET bottles, installed in a master inspection system (supplied as OEM via Krones as Aircontronic).
V&F
V&F MotoSense
Fast IMR-MS emission analyzer for engine and catalyst research, measuring hydrocarbons, nitrogen and sulfur compounds in exhaust gas with 20 ms response time.
V&F