V&F MotoSense

Manufacturer: V&F

V&F MotoSense

Fast IMR-MS emission analyzer for engine and catalyst research, measuring hydrocarbons, nitrogen and sulfur compounds in exhaust gas with 20 ms response time.

Key features

  • Response times of 20 ms for transient engine operation
  • Unlimited hydrocarbon measurement
  • NH3 and NOx measurement on SCR catalysts
  • Sulfur species for desulfation strategies

Technical data

Measuring principle
Ion-molecule reaction mass spectrometry (IMR-MS)
Mass range
7–160 amu
Resolution
< 1 amu
Analysis time
>= 100 ms
Measuring range
10^5
Response time
T90 < 150 ms
Detection limit
< 100 ppb
Concentration drift
< ±5 % over 24 h
Accuracy
< ±3 %
Precision
< ±3 %
Ambient temperature
20–35 °C
Max. humidity
80 % (non-condensing)
Gas consumption
30–3000 ml/min
Gas inlet temperature
50–190 °C
Gas inlet pressure
0.75–2 bar(a)
Power supply
230 V/50 Hz or 115 V/60 Hz, 800 W
Dimensions (W x H x D)
534 x 743 x 639 mm
Weight
75 kg

Description

Developing high-performance engines and catalytic converters with low fuel consumption and ultra-low emissions requires emission measurement systems that are very fast, sensitive and precise.

The V&F MotoSense gives engine and catalyst researchers a tool for investigating organic hydrocarbons and inorganic nitrogen or sulfur emissions in exhaust gas. Response times of 20 ms and a dynamic measuring range of 1 to 10^4 allow transient processes to be studied precisely.

Typical applications

  • Measurement of hydrocarbons (unlimited)
  • Monitoring of desulfation strategies in catalytic converters
  • Ammonia and NOx measurements on SCR catalysts

Application examples

Automotive industry

  • Ammonia and NOx measurements in engine exhaust gas
  • Desulfation strategies for catalytic converters

Measurable compounds

  • Nitrogen compounds: NH3, NO, NO2
  • Sulfur compounds: H2S, COS, SO2, CS2
  • Hydrocarbons: CH4, C2H2, C2H4, C3H6, C4H6 (butadiene), C6H6, C7H8, C8H8 (styrene), C8H10 (xylene), C4H8 (butylene), C5H10, C6H12 (cyclohexane), C7H14 (methylcyclohexane), C8H16, C9H18, C10H20

Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.

Ask about this machine