V&F HydrogenSense

Manufacturer: V&F

V&F HydrogenSense

Combined EI-MS and IMR-MS analyzer that measures all impurities in hydrogen for PEM fuel cells according to ISO 14687:2019 and SAE J2719.

Key features

  • Hydrogen purity testing to ISO 14687:2019 and SAE J2719
  • EI-MS and IMR-MS in one analyzer
  • IMR-MS detection limit 0.1 ppb
  • Covers sulfur, halogen and other critical impurities

Technical data

Measuring principle
IMR-MS and EI-MS combined in one analyzer
Mass range
7–519 amu (IMR-MS) / 1–100 amu (EI-MS)
Resolution
< 1 amu (IMR-MS / EI-MS)
Analysis time
100 ms (IMR-MS) / 32 ms (EI-MS)
Measuring range
10^5 (IMR-MS) / 0–100 vol% (EI-MS)
Response time
T90 < 1000 ms (IMR-MS / EI-MS)
Detection limit
0.1 ppb (IMR-MS) / 5 ppm (EI-MS)
Concentration drift
< ±5 % over 24 h
Accuracy
< ±3 %
Precision
< ±3 %
Ambient temperature
20–35 °C
Max. humidity
80 % (non-condensing)
Gas consumption
250–1000 ml/min
Gas inlet temperature
130 °C
Gas inlet pressure
1.5–5 bar(a)
Power supply
230 V/50 Hz or 115 V/60 Hz, 2000 W
Dimensions (W x H x D)
600 x 1692 x 700 mm
Weight
215 kg

Description

Hydrogen that contains impurities can damage fuel cells within a short time. The international standards ISO 14687:2019 and SAE J2719 therefore define hydrogen purity requirements and set maximum permissible levels for impurities such as H2O, O2, He, N2, Ar, CO, CO2, total sulfur, total hydrocarbons, HCHO, HCOOH, NH3, HBr, HCl and Cl2.

The V&F HydrogenSense combines an electron impact (EI-MS) and an ion-molecule reaction (IMR-MS) mass spectrometer in one analyzer unit to measure all impurities in hydrogen for PEM fuel cells.

Typical applications

  • Purity testing of hydrogen for PEM fuel cells

Measurable compounds

  • Permanent gases: H2, O2, N2, CO, CO2
  • Moisture: H2O
  • Noble gases: He, Ar
  • Hydrocarbons: CH4, C6H6, C7H8
  • Sulfur compounds: H2S, COS, CS2, R-SH (mercaptans)
  • Other: CH2O (formaldehyde), CH2O2 (formic acid), NH3
  • Halogenated compounds: HBr, HCl, Cl2, R-X (organic halides)

Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.