Manufacturer: V&F
V&F HydrogenSense
Combined EI-MS and IMR-MS analyzer that measures all impurities in hydrogen for PEM fuel cells according to ISO 14687:2019 and SAE J2719.
Key features
- Hydrogen purity testing to ISO 14687:2019 and SAE J2719
- EI-MS and IMR-MS in one analyzer
- IMR-MS detection limit 0.1 ppb
- Covers sulfur, halogen and other critical impurities
Technical data
- Measuring principle
- IMR-MS and EI-MS combined in one analyzer
- Mass range
- 7–519 amu (IMR-MS) / 1–100 amu (EI-MS)
- Resolution
- < 1 amu (IMR-MS / EI-MS)
- Analysis time
- 100 ms (IMR-MS) / 32 ms (EI-MS)
- Measuring range
- 10^5 (IMR-MS) / 0–100 vol% (EI-MS)
- Response time
- T90 < 1000 ms (IMR-MS / EI-MS)
- Detection limit
- 0.1 ppb (IMR-MS) / 5 ppm (EI-MS)
- Concentration drift
- < ±5 % over 24 h
- Accuracy
- < ±3 %
- Precision
- < ±3 %
- Ambient temperature
- 20–35 °C
- Max. humidity
- 80 % (non-condensing)
- Gas consumption
- 250–1000 ml/min
- Gas inlet temperature
- 130 °C
- Gas inlet pressure
- 1.5–5 bar(a)
- Power supply
- 230 V/50 Hz or 115 V/60 Hz, 2000 W
- Dimensions (W x H x D)
- 600 x 1692 x 700 mm
- Weight
- 215 kg
Description
Hydrogen that contains impurities can damage fuel cells within a short time. The international standards ISO 14687:2019 and SAE J2719 therefore define hydrogen purity requirements and set maximum permissible levels for impurities such as H2O, O2, He, N2, Ar, CO, CO2, total sulfur, total hydrocarbons, HCHO, HCOOH, NH3, HBr, HCl and Cl2.
The V&F HydrogenSense combines an electron impact (EI-MS) and an ion-molecule reaction (IMR-MS) mass spectrometer in one analyzer unit to measure all impurities in hydrogen for PEM fuel cells.
Typical applications
- Purity testing of hydrogen for PEM fuel cells
Measurable compounds
- Permanent gases: H2, O2, N2, CO, CO2
- Moisture: H2O
- Noble gases: He, Ar
- Hydrocarbons: CH4, C6H6, C7H8
- Sulfur compounds: H2S, COS, CS2, R-SH (mercaptans)
- Other: CH2O (formaldehyde), CH2O2 (formic acid), NH3
- Halogenated compounds: HBr, HCl, Cl2, R-X (organic halides)
Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.
Ask about this machine
Send us your question or the details of your project. We usually reply within one working day.
More in this range

V&F AirSense
Multi-purpose IMR-MS mass spectrometer for an unlimited number of gas components, built for flexibility and robustness across automotive, food, environmental and medical applications.
V&F
V&F CO2Sense
IMR-MS analyzer developed for measuring organic and inorganic impurities in CO2, meeting common food and beverage CO2 quality specifications such as ISBT and EIGA.
V&F
V&F PETSense
IMR-MS sniffer for the fast, accurate detection of contaminated refillable PET bottles, installed in a master inspection system (supplied as OEM via Krones as Aircontronic).
V&F