Manufacturer: V&F
V&F EISense
Classic quadrupole EI-MS analyzer with 70 eV ionisation and 1–100 amu mass range – the entry-level model for online multi-component gas analysis.
Key features
- Electron impact ionisation at 70 eV
- Measuring range 0–100 vol%
- Permanent gases, moisture, noble gases and light hydrocarbons
- Entry-level model for online multi-component analysis
Technical data
- Measuring principle
- Electron impact mass spectrometry (EI-MS), quadrupole
- Mass range
- 1–100 amu
- Resolution
- < 1 amu
- Analysis time
- >= 32 ms/amu
- Measuring range
- 0–100 vol%
- Response time
- T90 < 150 ms
- Detection limit
- 0.01 vol%
- Concentration drift
- < ±5 % over 24 h
- Accuracy
- < ±3 %
- Precision
- < ±3 %
- Ambient temperature
- 20–35 °C
- Max. humidity
- 80 % (non-condensing)
- Gas consumption
- 30–3000 ml/min
- Gas inlet temperature
- 50–190 °C, adjustable
- Gas inlet pressure
- 0.75–2 bar(a)
- Power supply
- 230 V/50 Hz or 115 V/60 Hz, 800 W
- Dimensions (W x H x D)
- 534 x 743 x 639 mm
- Weight
- 61 kg
Description
The V&F EISense is a classic quadrupole mass spectrometer with electron impact ionisation (EI-MS) and is the entry-level V&F model for online multi-component gas analysis.
Ionisation takes place at 70 eV; the mass range is 1 to 100 amu with a resolution of 1 amu.
Typical applications
- Process monitoring of combustion plants
- Development of combustion engines (diesel and petrol)
- Alternative fuels
- Catalyst development
- Hydrogen measurements
Application examples
Automotive industry
- Hydrogen measurements in combustion engines
- Gas analysis in fuel cell technology
Steel industry
- Process control in steel production
- Monitoring of hazardous exhaust gases in steel production
- Hydrogen production in a cyclic steelmaking process
Environmental industry
- Online quantification of tar during biomass gasification
Polymer, tyre and rubber industry
- Evaporation of inorganic and organic compounds from elastomers
Measurable compounds
- Permanent gases: H2, O2, N2, CO, CO2
- Moisture: H2O
- Noble gases: He, Ne, Ar, Kr, Xe
- Hydrocarbons: CH4, C6H6, C7H8
Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.
Ask about this machine
Send us your question or the details of your project. We usually reply within one working day.
More in this range

V&F AirSense
Multi-purpose IMR-MS mass spectrometer for an unlimited number of gas components, built for flexibility and robustness across automotive, food, environmental and medical applications.
V&F
V&F CO2Sense
IMR-MS analyzer developed for measuring organic and inorganic impurities in CO2, meeting common food and beverage CO2 quality specifications such as ISBT and EIGA.
V&F
V&F PETSense
IMR-MS sniffer for the fast, accurate detection of contaminated refillable PET bottles, installed in a master inspection system (supplied as OEM via Krones as Aircontronic).
V&F