Skip to content
Green Fields Instruments
Menu
V&F EISense

Manufacturer: V&F

V&F EISense

Classic quadrupole EI-MS analyzer with 70 eV ionisation and 1–100 amu mass range – the entry-level model for online multi-component gas analysis.

Key features

  • Electron impact ionisation at 70 eV
  • Measuring range 0–100 vol%
  • Permanent gases, moisture, noble gases and light hydrocarbons
  • Entry-level model for online multi-component analysis

Technical data

Measuring principle
Electron impact mass spectrometry (EI-MS), quadrupole
Mass range
1–100 amu
Resolution
< 1 amu
Analysis time
>= 32 ms/amu
Measuring range
0–100 vol%
Response time
T90 < 150 ms
Detection limit
0.01 vol%
Concentration drift
< ±5 % over 24 h
Accuracy
< ±3 %
Precision
< ±3 %
Ambient temperature
20–35 °C
Max. humidity
80 % (non-condensing)
Gas consumption
30–3000 ml/min
Gas inlet temperature
50–190 °C, adjustable
Gas inlet pressure
0.75–2 bar(a)
Power supply
230 V/50 Hz or 115 V/60 Hz, 800 W
Dimensions (W x H x D)
534 x 743 x 639 mm
Weight
61 kg

Description

The V&F EISense is a classic quadrupole mass spectrometer with electron impact ionisation (EI-MS) and is the entry-level V&F model for online multi-component gas analysis.

Ionisation takes place at 70 eV; the mass range is 1 to 100 amu with a resolution of 1 amu.

Typical applications

  • Process monitoring of combustion plants
  • Development of combustion engines (diesel and petrol)
  • Alternative fuels
  • Catalyst development
  • Hydrogen measurements

Application examples

Automotive industry

  • Hydrogen measurements in combustion engines
  • Gas analysis in fuel cell technology

Steel industry

  • Process control in steel production
  • Monitoring of hazardous exhaust gases in steel production
  • Hydrogen production in a cyclic steelmaking process

Environmental industry

  • Online quantification of tar during biomass gasification

Polymer, tyre and rubber industry

  • Evaporation of inorganic and organic compounds from elastomers

Measurable compounds

  • Permanent gases: H2, O2, N2, CO, CO2
  • Moisture: H2O
  • Noble gases: He, Ne, Ar, Kr, Xe
  • Hydrocarbons: CH4, C6H6, C7H8

Values marked as application-dependent (measuring range, response time, detection limit, gas consumption) depend on the compounds measured, the system configuration and the instrument settings.