Manufacturer: V&F
V&F GasOxidizer
Software-controlled tube furnace that oxidises all sulfur components in engine exhaust gas to SO2 for fast, time-resolved online oil consumption measurement with IMR-MS.
Key features
- Oxidises all sulfur components to SO2
- Furnace temperature up to 1200 °C
- Software-controlled operation
- Used with the V&F LubeSense for oil consumption measurement
Technical data
- Max. temperature
- 1200 °C
- Warm-up time
- 1 h
- Cooling time
- approx. 5 h
- Ambient temperature
- 20–35 °C
- Gas consumption
- 1 l/min
- Max. gas inlet pressure
- 6 bar(a)
- Oxygen consumption
- 80 ml/min
- Power supply
- 230 V/50 Hz, 1200 W
- Dimensions (W x H x D)
- 860 x 490 x 285 mm
- Weight
- 24 kg
Description
The V&F GasOxidizer is a software-controlled tube furnace for oxidising the exhaust gas of combustion engines. It was developed to provide fast, time-resolved online measurement of engine oil consumption.
The method uses sulfur contained in the oil as a tracer. The sulfur passes through a chemical conversion in the combustion chamber, is determined in the exhaust gas and correlated with the oil consumed. In the SO2 tracer method the sulfur content is obtained by measuring sulfur dioxide (SO2) in the exhaust gas.
The GasOxidizer ensures that all sulfur components in the exhaust gas are measured as SO2: oxygen (O2) is added and converted to ozone (O3), which oxidises the sulfur species to SO2. SO2 is then quantified by ion-molecule reaction mass spectrometry (IMR-MS). This shortens measuring times and makes the low oil consumption of modern engines measurable.
Typical applications
- Oil consumption measurements
- Measurement of total sulfur
Application examples
Automotive industry
- Oil consumption measurements on combustion engines
Ask about this machine
Send us your question or the details of your project. We usually reply within one working day.
More in this range

V&F LubeSampler
Continuous engine oil sampling from the oil sump with fuel separation, for online oil dilution measurement with the V&F FuelSense.
V&F
V&F AutoSampler
Headspace autosampler based on the Agilent G1888 for automatic analysis of gas samples in sealed vials with V&F mass spectrometers – 70 vials, 12 oven positions.
V&F
V&F PreSampler
Heated sample preparation module for engine exhaust gas, with heated pump, inlet pressure regulator (0.8–5 bar abs.) and 0.5 µm ceramic particle filter with backflush.
V&F